Total Hip Arthroplasty Implant (ISO 7206-4): Ti-6Al-4V Stem & CoCrMo Head
Two-body orthopedic fatigue-test assembly (Ti-6Al-4V ELI femoral stem + CoCrMo Ø32 mm modular head joined across a 12/14 Morse taper via Bonded MPC) under a 2,300 N vertical hip joint reaction force applied through an RBE2 spider at the head center.
ISO 7206-4 specifies the endurance test setup for total hip joint prostheses: the distal femoral stem is potted in bone cement while a vertical test load F_z = −2,300 N is applied through the center of the modular femoral head at a 30 mm medial offset (bending moment M = 69 N·m). This 2-body CAD assembly models the Ti-6Al-4V ELI stem (E = 114 GPa, ν = 0.34) and CoCrMo Ø32 mm head (E = 220 GPa, ν = 0.30) coupled across their conical Morse taper via Bonded MPC, with an RBE2 spider transferring F_z = −2,300 N from the head center (30, 0, 130) mm onto the upper bearing hemisphere.
1. Geometría y malla FEA
- •2 B-Rep solid bodies: Body 1 — Ti-6Al-4V ELI femoral stem with distal cement fixation zone, curved transition, Ø14 mm neck, and Morse taper trunnion; Body 2 — CoCrMo Ø32 mm modular femoral head.
- •Region-specific materials: Ti-6Al-4V ELI (E = 114,000 MPa, ν = 0.34, ρ = 4,430 kg/m³) on Body 1 and ASTM F1537 CoCrMo (E = 220,000 MPa, ν = 0.30, ρ = 8,300 kg/m³) on Body 2.
- •2nd-order curved tetrahedral mesh (P2 / TET10): 1,861 elements, 3,077 nodes, and 9,231 DOFs (single-step linear static solve without multi-step slider).
2. Cargas y condiciones de contorno
- •Fixed support (Ux = Uy = Uz = 0) on the distal potted stem surfaces (surfaces 1, 2, 3 up to z = 50 mm).
- •Bonded (MPC) contact pair across the conical Morse taper interface between Body 1 and Body 2.
- •Remote Force (RBE2 spider) F_z = −2,300 N at reference point (30, 0, 130) mm coupled to the upper bearing hemisphere of the CoCrMo head (surface 8).
3. Resultados de simulación y conclusiones
- ✓Peak von Mises stress σ_vm = 180.72 MPa at the lateral neck–stem transition (within 0.14% of combined Euler–Bernoulli bending + axial compression theory σ_nom = 180.47 MPa and 4.3× below the 795 MPa yield strength of Ti-6Al-4V ELI).
- ✓Maximum head displacement |U|_max = 0.1942 mm at the CoCrMo femoral head.
- ✓Exact vertical reaction equilibrium R_z = +2,300.0 N (0.00% error) and overturning moment M_y = 69.00 N·m.
WebCAE enables medical device engineers to run ISO 7206-4 bi-metallic hip implant stress and stiffness checks with true CAD geometry, Morse taper MPC coupling, and RBE2 spherical head loading in seconds.
14 días de acceso completo a Pro. El primer cobro se realiza tras 14 días salvo cancelación previa.
- ✓Prueba de 14 días ($0 hoy) (cancela cuando quieras)
- ✓Hasta 1,5M+ GDL (solucionador disperso directo)
- ✓Apps nativas para Windows, macOS y Linux (100% offline)
- ✓Contacto no lineal y pandeo (Kσ, MITC4, RBE2/RBE3)
Pregunta a un ingeniero sobre tu modelo
Describe tu pieza, caso de carga o formato CAD — responderemos en 2 horas laborables.
¿Quieres una respuesta instantánea sobre teoría FEA o este modelo?
Haz clic en el chat de IA en la esquina inferior derecha para consultar sobre condiciones de contorno, criterios de tensión, convergencia de malla o scripts DSL.